It is suitable for a series of tests on color fastness and color fastness to airborne contaminants, such as knitted fabrics, woven fabrics, synthetic fiber fabrics, natural fiber fabrics, etc. It can also be used for the ozone color fastness test of fabrics such as nylon carpet.
Standards
AATCC 109-2011 Colorfastness to Ozone in the Atmosphere under Low Humidities.
AATCC 129-2011 Colorfastness To Ozone In The Atmosphere Under High Humidities
ISO 105-G03-1993 Textiles; tests for color fastness; part G03: color fastness to ozone in the atmosphere
GB/T 11039.3-2005 Textiles; tests for color fastness; part 3: color fastness to ozone in the atmosphere
JIS L0890 Test methods for color fastness to ozone
Features
· Multi-functional rotating sample rack supports hanging and placing samples in various ways to meet the test requirements of samples in different shapes and specifications. Sample rack speed is adjustable.
· High output ozone generator with closed-loop ozone concentration control system and easy parameter setting panel provides a powerful control function and accurate test results.
· Transparent window with internal lighting system provides convenience to observe the test process. The security door lock ensures test safety.
Structure
1. The apparatus consists of an ozone generator, fans, baffles, sample holder and the sample compartment composition. The use of high-voltage transformer ozone generator discharge through the grid, the generator is placed in front of the low pressure fan, an ozone separator system injects ozone into the test chamber. Sample holder mounted on the wire can be applied to samples of various sizes.
2. The device is equipped with high-efficiency ozone generator, controller, closed-loop control of ozone concentration and can accommodate 24 rotating sample holders. The pre-set parameters can be displayed on the screen , thereby increasing the accuracy of the device operation and test results.
Technical Parameters
Items | Parameters |
Temperature Range | 10℃-50℃ |
Temp Accuracy | ±1.0℃ |
Humidity range | 50-90% RH |
Humidity accuracy | ±3% |
Ozone concentration range | 0-500pphm |
Adjust method | flow regulation |
Ozone generation | Static air emissions |
Internal dimension | 400mm x 400mm x500mm |
Power supply | 220V,50Hz or115,60Hz |
External dimension | 1000mm×650 mm×1450 mm |
Weight | 95KG |