II.Main technical
features
1.Designed as per GB/T 5208-2008
Determination of flash point - Rapid equilibrium closed cup method. It can
realize purpose of testing closed cup flash point rapidly and at low
temperature.
2.The instrument adopts semiconductor refrigeration device and
external water cooling device.The size is small and cooling rate is rapid.Easy
to operate.
3.Test sample needed is less.2ml for each time(4ml for solid
sample and semi-solid sample)
4.The determination procedure is automatically
completed except ignition.Test results can be printed automatically.
III.Main technical specifications and parameters
1.
Power supply: AC(220±10%)V, 50Hz
2. Flash point detecting range:
(50~100)°C
3. Precision of determination:
Absolute difference value
between two determinations is lower than 2°C(same operator)
Absolute
difference value between two determinations is lower than 3°C(different
operator) 4. Accuracy of
temperature control: ±0.5°C
5. Igniting device:Electric igniting gun
6.
Cooling mode:Semiconductor(with cold water cycle at external
connection)
7. Overall Power Consumption:Not higher
than 300W
8. Ambient temperature:(5~30)°C
9. Relative
humidity:(30~80)%
10. Overall
dimension:370mm×280mm×280mm(L×W×H)
Notification: Our company can customize
instruments with different ranges of low-temperature flash point according to
customer's need.