PN-2000FB Respiratory Humidifier (Support Heated Wire)

Minimum Order
1
Packaging
N/A
Delivery
15 Days
Pigeon Medical Respiratory Humidifier PN-2000FB offers an integrated solution to provide humidified and warmed gas to those patients in need of mechanical ventilation or other respiratory supports. With built in servo control feedback system and alarm system, it incorporates to monitor constantly and provide visible and audible alarm according to the proximal and distal temperature respectively.

In addition, the over-heat protection will automatically perform thermal cut-off once the temperature of the heating plate excess 95oC to avoid any unexpected circumstances. PN-2000FB also supports heated wire on both disposable and reusable breathing circuit, ensure to avoid temperature loss and water condensation while the medical gas travels through the circuit.

Working Principle of PN-2000FB Respiratory Humidifier

Pigeon Medical Respiratory Medical Humidifier PN-2000FB is designed to use in hospital environment which can be capable with mechanical/ non-mechanical ventilation and oxygen therapy for adults. Meanwhile, it supports invasive ventilation, infant resuscitation and bubble CPAP and oxygen therapy for neonatal care.

With optional heated wire function, it matches with both disposable/reusable and heated wire/non-heated wire breathing circuit.

Features of PN-2000FB Respiratory Humidifier

Heated or non-heated wire temperature control with 3 different phases independently

Match with all kind of ventilators

Provide different connectors for pediatrics and adults breathing circuit

Built-in alarm function

Provide high quality aluminium clamp and slot plate for the blender to safely hang on the trolley
Specification of PN-2000FB Respiratory Humidifier
Model PN-2000FB
Power Source AC 220V, 50-60Hz (Optional: AC 110V, 50-60Hz)
Heated Plate 150W
Thermal Cut Off 95±6℃
Display Three digit, 0.56 inch, 7 segment Led
Temperature Monitoring 0-75℃
Max-temperature of Heated Plate 85℃
Heated Wire AC 23V 60W, 50-60Hz
Power Rate 220W
Alarm Yes
Temperature Control Setting Heated Wired Mode:
P1: 25-29℃, P2: 29-35℃,
P3: 35-39℃
Non-heated Wired Mode:
P1: 23-27℃, P2: 27-32℃,
P3: 32-37℃
Range 10-70oC, Accuracy: ±2℃ (In 25-45℃ temperature range)

  • Country: China (Mainland)
  • Business Type: Manufacturer
  • Market:
  • Founded Year:2001
  • Address:No.118,Block 6,ZhuYuan,ChiWei Township,PuNing City, GuangDong,China.
  • Contact:pigeon medical
more images of PN-2000FB Respiratory Humidifier (Support Heated Wire)
*Your name:
*Your Email:
*To:Guangdong Pigeon Medical Apparatus Co., Ltd
*Subject:
*Message:
Enter between 20 to 3,000 characters. English only.     Characters : 0 / 3000
*Enter the secure code shown below Mfrbee security Image      Reload Image

submiting now We do inquire for you , please wait ...

B-BRAND ATLAS BACK SUPPORT MED B-Brand FILTERS Other Respiratory Range

Jays Engineering Supply - B-Brand Other Range. Simple Economy Belt 4"/10cm Wide. Made from Rubber 'Bio Gel' with LECRA and Polyester Covering. Simple One Action Velcro and Buckle Fastening. Individually Packaged with Header Card. ...

Jays Engineering Supplies Ltd.