Main Features:
A lot of new technology and new materials are adopted. Such features as LCD display, menu-type instruction operation, curve storage, flameout protector and outside printout are enabled by the application of top grade embedded mono-chip computer circuit
Specifications:
| Linear Error : | K: ≤0.005mmol/L;Na: ≤0.03mmol/L |
| Related coefficient | bigger than 0.999 |
| Measuring range | K: 0~10mmo|/L, Na: 0~200mmo|/L |
| Reproducibility | (GV%):K: 1.5%(5mmo|/L or 50mmo|/L) , Na: 1.0%(140 mmol /L) |
| Linear error: Response time | K: 2%; Na: 2%: not longer than 8s |
| Solution consumption | 4ml /min; flameout detection, automatic gas cut-off function |
Working conditions:
Environmental Temperature: 10~35°C
Relative humidity: ≤85%
The instrument should be placed horizontally, avoiding direct sunshine radiation. There is no disturbance of strong electromagnetic field or vibration
Fire extinguishing device should be equipped in the application field and good aeration should be guaranteed. Power supply: AC220V±22V, frequency 50Hz±1Hz,with good grounding
Standard configuration:
Main Instrument 1
Power Cable 1
Chimney 1
Clearing rod 2
Air compressor 1
Printer SP-T 2425SK 1
Optional accessories:
Printer SP-T 2425SK
Air compressor
Lithium filter
Overall dimension & weight:
Shipping package dimension: 810*610*430mm
Gross weight: 23Kg