Industrial Free chlorine analyzer POP-2200

POP-2200 Integrated Free Chlorine /CLO2/pH Online Analyzer comes together intelligent control meter, Free Chlorine/CLO2sensor, pH sensor, Temp probe and matching constant flow device and other relative accessories,with tight structure, easy transportation and installation and free of maintenance features.

Characteristics:

a.Wall mounted integrated Free Chlorine/CLO2 online analysis system integration;

b.Embedded free chlorine, chlorine dioxide analysis software of two mathematical models;

c.3.5\"LCD color display,bilingual Chinese/English ,complete operation guided ;

d.Constant flow rate control with the patents, make sure the measurement accuracy,do not changed by

the the pipeline pressure ;

e.Anti-siphon design, the system automatically in the conservation of the state after the shutdown;

f.Double channels isolated (4~20)mA,Instrument / Transmitter for selection;

g.High / low limit setting relay*2,Photoelectric programmable*1, switching / pulse regulating

state;

h.Protective plastic chassis, preset back cover, the independent panel.

Technical Parameters:

Model

POP-2200(CFA-2200)

POP-2200(CRA-2200)

Sensor

Free Chlorine/pH/Temp/

CLO2/Temp

Measurement range

Free Chlorine

(0.10~2.00)mg/L(ppm);(0.10~20.00)mg/L(ppm)

CLO2

(0.10~2.00)mg/L(ppm);(0.10~20.00)mg/L(ppm)

pH

2.00~12.00(for the Free Chlorine)

Temp

(0.0~99.9)℃(Temp. Compensation Pt1000)

Resolution

Free Chlorine

0.01mg/L

CLO2

0.01mg/L

pH

0.01

Temp

0.1℃

Accuracy

Free Chlorine

90% confidence interval ±10%FS

CLO2

90% confidence interval ±10%FS

pH

0.1level

Temp

±0.5℃

Medium

pH

6.0~8.0(only for Free Chlorine)

Temp

(0.0~40.0)℃

(4~20)mA

Output

Channel

Double

Technical feature

Isolated,Reversible, fully adjustable, instrument/transmitter mode for selection

Channel configuration

Programmable for free chlorine, CLO2, Temp, pH

Loop resistance

400Ω(Max),DC 24V

Accuracy

±0.1mA

Control port

channels

Triple channels

Contact mode

Photoelectric semiconductor relay

Load capacity

Load current 50mA(max),AC/DC 30V

Functions selection

Programmable( free chlorine, CLO2, Temp, pH, timing) output;

Communication port

RS485

MODBUS RTU RS485

Working power

Power supply voltage

System AC 220V /50Hz, indicator DC24V.

Power consumption

≤5.5W

Working environment

temperature:(0~50 )℃;humidity:≤85% RH(none condensation)

Storage environment

Temperature:(0~60)℃; humidity:≤85%RH(none condensation)

Protection level

IP54(plastic protection cabinet, integration system)

Cabinet material

ABS

Cabinet size

380mm×280mm×130mm(H*W*D)

installation

Wall mounted(with the preset back cover)

Cabinet weight

≤10kg

Pipeline flow

(250-350)mL/min

  • Company: Hebei Bossen Trading Co.,Ltd.
  • Country: China (Mainland)
  • Business Type: Trading Company,Manufacturer
  • Market: Asia,Europe
  • Founded Year: 2003
  • Address: No.8 Huifeng Road, Qiaoxi Dist.,Shijiazhuang China
  • Contact: David Lee

Hebei Bossen Trading Co.,Ltd.

Enter between 20 to 3,000 characters. English only. Characters: 0 / 3000
submiting now We do inquire for you, please wait...