KAPTA 2000 pH/ ORP Free Active Chlorine Temperature Measuring & Control Instruments

Minimum Order
1
Brand Name
Endetec
Place Of Origin
France
Packaging
N/A
Delivery
15 Days

pH/ ORP, Free Active Chlorine, Temperature Measuring & Control Instruments
chlorine (HOCl, HOBr or ClO2 ) & temperature


KAPTA 2000 Water Quality Controller
The KAPTA 2000 controller has been designed specifically for the industrial and water treatment markets and, coupled with the KAPTA 2000-AC2


The controller is equipped with all the key functions required to provide the user with flexibility for set-up, measurement and control of systems from simple chlorine measurement and recording right through to industrial process control and chemical dosing.


The KAPTA 2000 controller is capable of measuring chlorine, ORP, pH and temperature and is fitted with 2 current outputs, 4 relays and P.I.D. functionality allowing up to 6 external points to be measured or controlled. Operation of the controller is simple with a 128x64 pixel resolution graphic display providing continuous readout of chlorine / temperature measurement and status of the control output devices, selectable in a choice of 5 different languages (English, French, German, Italian and Spanish).


The controller is IP66 rated, wall mounted and equipped for RS485 serial port communication for remote monitoring and data storage. On-board software is designed for 2 point calibrations to ensure accuracy, reliable performance and sensor diagnostic information.


KAPTA 2000 Water Quality Controller Benefits
• Real-time water network monitoring
• No on-going membrane / electrolyte maintenance
• Chemical free operation


KAPTA 2000 Water Quality Controller Main features
• Amperometric sensor
• Measurement of the main disinfection species: HOCl, HOBr or ClO2
• User replaceable sensing head
• Inline monitoring at up to 10 bar pressure
• Easy conditioning
• 1 year shelf life


KAPTA 2000 Water Quality Controller Typical applications
• Water quality control
• Drinking water security
• Industrial process control
• Chemical dosing measurement and control
• RO membrane security
• Swimming pool water quality
• Cooling towers
• Fish farming


KAPTA 2000 Water Quality Controller Specifications (Controller)

General specifications
• Measurement of free chlorine or active chlorine (HOCl), temperature (PT 1000) and pH (optional)
• Simultaneous display of chlorine and temperature
• Wall mounted
• Internal flash 4 Mbit memory
• PID adjustment and control
• Serial output RS485 modbus RTU
• 2 Programmable analogue outputs
Display
Back-lit LCD 128x64 pixel
Sensor Inputs

Analogue Inputs
Measurement and Measurement /
Control
Digital Communication
Serial RS485 Modbus RTU
4 - 20mA Outputs

Relay Outputs

Measurement accuracy
±2.5% F.S. (chlorine)
±1.0% F.S. (temperature)
Dimensions
144mm x 144mm x 122.5mm
Operating temperature range
-10 to +130 degrees C
Electrical Rating
IP66
Weight
1 Kg
Power Consumption
< 7 W
Enclosure material
ABS, Grey RAL 745 (body)
Polycarbonate, UV resistant (Frontal)
Power Rating
100 to 240 Vac
  • Country: Hong Kong
  • Business Type:
  • Market:Europe
  • Founded Year:2014
  • Address:
  • Contact:Rita yu
more images of KAPTA 2000 pH/ ORP Free Active Chlorine Temperature Measuring & Control Instruments
*Your name:
*Your Email:
*To:ISweek company
*Subject:
*Message:
Enter between 20 to 3,000 characters. English only.     Characters : 0 / 3000
*Enter the secure code shown below Mfrbee security Image      Reload Image

submiting now We do inquire for you , please wait ...