A:Brief description:
DSHD-261-1 Pensky-Martens Closed Cup Flash Point Tester is manufactured as per standards GB/T 261-2008 Determination of Flash Point(Pensky-Martens Closed Cup) and ASTM D93.
DSHD-261-1 Pensky-Martens Closed Cup Flash Point Tester issuitable to make determination of the closed cup flash point of the petroleum products which flash point over 40°C.
B:Main advantages:
1. Single chip microcomputer control technique. LCD screen displays,English man-machine interaction interface. The heating power is continuously step-less adjustable. The rate of heating up is exact,Easy setting for parameters.
2. The LCD screen has prompting menu, prompting input operation interface. It shows the setted parameters, real-time temperatures and other test parameters. Press the"record" button when the flash point appears. The screen will show and reserve the flash point.
3. The structure is newly design and small. Stainless steel table-board looks clean and beautiful. It can be used as long as equipped with gas or other civilian fuels. It is easy to use and test results exact.
C: Main technical specifications and parameters
1.Power supply:AC( 220±10% )V, 50Hz.
2.Heating device:(1) The furnace is silicon carbide material.Power is 600W heating wire.
(2)The heating power is adjustable from 0 W to 600W.
3.Rate of sample's heating up:Procedure A: (5-8 )°C/min.
Procedure B :(1-1.5)°C/min.
Automatically control
4.Electric stirring device:(1)Stirring motor:(BYGH101stepping motor)
(2) Driving mode: flexible shaft
(3) Stirring vane size: 8mm×40mm
5.Stirring rate:Accord with the requirements of GB/T 261-2008.
(1) (90-120 ) turns/min,suitable to test procedure A.
(2)(250±10 ) turns/min,suitable to test procedure B.
6.Igniting device: (1) Igniting source: gas(or other civilian fuels,the same below)
(2) Bore diameter: 0.8mm
7.Temperature testing probe:PT100
8.Ambient temperature: no more than 35°C
9.Relative humidity: no more than 85%
10.Power consumption of overall unit: no more than 650W
A:Brief description: DSHD-5208 Rapid Closed Cup Flash Point Tester is manufactured as per GB/T 5208-2008 Determination of flash point - Rapid equilibrium closed cup method. DSHD-5208 Rapid Closed Cup Flash Point Tester Conforms tothe requirements of rapid ...
A:Brief description: DSHD-261A Automatic Pensky-Martens Closed Cup Flash Point Tester is designed according industrial standard SH/T0315 Technical Specifications of Closed Cup Flash Point Tester. The main function ofDSHD-261A Automatic Pensky-Martens Closed Cup Flash ...
A:Brief description: DSHD-261D Fully-automatic Pensky-Martens Closed Cup Flash Point Tester is designed accordingthe standards GB/T261-2008 Test Method for Flash Point of Petroleum Products by the Pensky-Martens Closed Cup Method and ASTM D93. Its function isto ...
A:Brief description: DSHD-3536A Automatic Cleveland Open Cup Flash point tester is designed according the standards GB/T3536-2008 Test Method for Flash Point and Fire Point of Petroleum Products - Cleveland Open Cup Method, ASTM D92 and ISO 2592. DSHD-3536A Automatic ...
A:Brief description: DSHD-3536-1 Cleveland Open-Cup Flash Point Tester is designed according the standards GB/T3536-2008 Test Method for Flash Point and Fire Point of Petroleum Products - Cleveland Open Cup Method,ASTM D92 and ISO2592. DSHD-3536-1 Cleveland Open-Cup Flash ...
A:Brief description: DSHD-3536 Cleveland Open-Cup Flash Point Tester is designed according the standards GB/T3536-2008 Test Method for Flash Point and Fire Point of Petroleum Products - Cleveland Open Cup Method,ASTM D92,ISO2592. DSHD-3536 Cleveland Open-Cup Flash Point ...
A:Brief description: DSHD-267 Open Cup Flash Point Tester is designed accordingthe National Standard GB/T267\"Test Methods for Flash Point and Fire Point of Petroleum Products (Open Cup Methods)\" and Industry Standard of Petrochemical SH/T0318\"Technical Requirements for ...
Power supply AC(220±10%)V, 50Hz Temperature measurement Measuring range: Room temperature~350°C Test range of flash point: 41°C~350°C Flash point≤104°C: error ≤2°C Flash point>104°C: error ≤6°C Accuracy: 0.1°C Heating rate Procedure A: (5~6)°C/min Procedure ...
Ⅰ.Introduction GD-3536D ASTM D92 Fully-Automatic cleveland Open Cup Flash Point Tester is designed and made as per the National Standard of People’s Republic of China GB/T3536 “Test Methods for Flash Point and Fire Point of Petroleum Products (Cleveland Open Cup ...